Baxter Recalls TruSystem Operating Tables Due to Unexpected Lift Mechanism Drops
Baxter Healthcare is recalling select TruSystem operating tables because the main lift mechanism can unexpectedly drop during surgery.
Baxter Healthcare Corporation is recalling select models of TruSystem operating tables and table columns, including the TruSystem 7500 Hybrid Plus (FC), TruSystem 7000 U (dV), TS7500 MOBIUS, TruSystem 7000 U, and TruSystem 7500 surgical table.
The recall was initiated because the main lift assembly, or telescopic spindle, can experience unexpected wear. This can cause the tabletop to unexpectedly drop to its lowest position during use, and the table cannot be raised after dropping. An uncontrolled drop during surgery may result in abrupt patient displacement, posing risks of traumatic injury, airway compromise, and surgical complications. Baxter has received three complaints related to the issue, with no reported injuries.
The affected surgical tables were distributed worldwide and nationwide in the United States. The recall is classified by the U.S. Food and Drug Administration (FDA) as a Class II recall.
What to do
- Stop using the affected operating tables and columns immediately.
- Contact Baxter Healthcare Corporation for further instructions regarding the recalled equipment.
- Report any equipment failures or adverse events to Baxter or the FDA.
At a glance
| Agency | U.S. Food and Drug Administration (FDA) |
|---|---|
| Recall numbers | Z-0354-2027, Z-0353-2027, Z-0351-2027, Z-0352-2027, Z-0350-2027 |
| Date | 2026-11-11 |
| Company | Baxter Healthcare Corporation, Deerfield, IL |
| Severity | Class II |
| Quantity | 1 unit; 62 units; 77 units; 16 units; 129 units |
| Where sold | Worldwide – US Nationwide distribution. |
| Lot codes / identifiers | Serial numbers: 107971925; UDI-DI number: 00887761974067; Serial numbers: 108920104, 110192607, 110192885, 110208936, 110212885, 110220066, 110227520, 110248337, 110250614, 109423464, 110192608, 110192886, 110208941, 110212886, 110220068, 110227523, 110248343, 110250621, 109517138, 110192609, 110206656, 110210717, 110220055, 110220069, 110227524, 110249706, 110250622, 110139767, 110192683, 110206657, 110210718, 110220056, 110220071, 110227525, 110249712, 110206662, 110139776, 110192882, 110206659, 110210720, 110220057, 110220072, 110227552, 110250605, 110188921, 110188919, 110192883, 110206660, 110210721, 110220058, 110220078, 110227553, 110250609, 110188922, 110188920, 110192884, 110206664, 110212881, 110220065, 110220079, 110248218, 110250613; UDI-DI number: 00887761967519; Serial numbers: 108094281, 108468075, 108484417, 108506318, 108516636, 108536785, 108544115, 108604154, 108463223, 108484404, 108489634, 108511411, 108516637, 108544106, 108544116, 108618964, 108463224, 108484405, 108489648, 108511412, 108521436, 108544107, 108604137, 108618965, 108468062, 108484406, 108506316, 108511413, 108521452, 108544110, 108604140, 108468063, 108484408, 108506317, 108516634, 108536776, 108544111, 108604153; UDI-DI number: 00887761974425; Serial numbers: 108517025, 110115554, 110144247, 110188916, 109558596, 110139748, 110166201, 110188918, 109737582, 110139769, 110177678, 110192686, 110115551, 110144245, 110185616, 109889060; UDI-DI number: 00887761967526; Serial numbers: 108491706, 108726496, 108758657, 108858029, 108945825, 109083212, 109343910, 109684319, 110016886, 108557062, 108736192, 108773032, 108882327, 108976094, 109174333, 109370104, 109690845, 110016887, 108574826, 108736193, 108773033, 108882328, 108976095, 109174334, 109396256, 109690846, 110072583, 108578755, 108736194, 108842519, 108883564, 108976101, 109174335, 109464647, 109708813, 110084982, 108580717, 108736195, 108842523, 108883565, 109047445, 109174336, 109464648, 109708816, 110086138, 108580718, 108743656, 108842524, 108896687, 109047446, 109228237, 109464649, 109885369, 110144223, 108580720, 108743657, 108842525, 108896688, 109047447, 109228238, 109464650, 109896735, 110144224, 108604827, 108745163, 108842526, 108912913, 109047448, 109249973, 109488031, 109896736, 110144225, 108604828, 108745164, 108842679, 108912914, 109047451, 109249974, 109493243, 109896737, 110158878, 108604832, 108745165, 108842680, 108931731, 109047452, 109249989, 109500155, 109896738, 110158879, 108604833, 108745166, 108849690, 108931732, 109047460, 109301944, 109561673, 109933665, 110175899, 108726492, 108745180, 108849691, 108945772, 109047461, 109301945, 109561674, 109948691, 110175900, 108726493, 108745181, 108849694, 108945823, 109078531, 109301946, 109571317, 110007863, 110163294, 108726495, 108758656, 108849695, 108945824, 109078533, 109343482, 109571318, 110016883, 110163293, 110183861, 110220396, 110220395; UDI-DI number: 00887761968639 |
Recalled products
- TruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087
- Operating Table TruSystem 7000 U (dV), Model/Catalog Number: 1723633
- Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695
- Operating Table TruSystem 7000 U, Surgical table, universal, Model/Catalog Number: 1604788
- TruSystem 7500 surgical table, Model/Catalog Number: 4091000
Source: official FDA notice. This summary was written by 365 News from the official notice. Details can change; check the notice before acting.